AFC Module MeetingFriday - May 21, 2004
Report on the Window burst test
Edgar Black
IIT
MTA Absorber Design Progress
This windows are of a nominal 21cm diameter
The diameter was selected to fit on the forced flowabsorber inside the 40cm. diameter magnet bore(lab G magnet)
The design called for 25psi MOP and rupture up to100psi
The bellow type new configuration was aimed toachieve a minimum of 105 micron thickness at thecenter
Data from the burst test on this window will beused to verify the window design parameters forMICE
Fabricated window design details
Window profile parameters
Manifold construction detail
Back plate for pressure test
Receiving window inspection CMC report (partial sample)
the holes and the plane on right side.
Bolt circle constructed of points of intersection between
DIM BOLT CIRCLE OF HOLES ON RT SIDE= LOCATION OF CIRCLE B.C.RT SIDE  UNITS=MM
AX    MEAS
X     0.00000
Y     0.00000
Z     0.00000
D   218.02053
RN    0.04067
 
Holes measured as cylinders, starting with the hole
at 90 deg and going anti-clockwise. Measurements shown
are  Polar Radius, Polar Angle, Diameter and  Vectors I,J,K .
DIM HOLE AT 90 DEG= LOCATION OF CYLINDER RT 90 HOLE  UNITS=MM
AX    MEAS
PR  108.98744
PA   89.98625
D     3.21533
V   -0.00063 -0.00067 1.00000    0.00000
 
DIM RND1= ROUNDNESS OF CYLINDER RT 90 HOLE  UNITS=MM
AX   MEAS
M   0.10550
 
DIM HOLE AT 150 DEG= LOCATION OF CYLINDER CYL1  UNITS=MM
AX    MEAS
Measurements results
Photogrametry technique give the minimumthickness to be 190 microns
Gauged thickness on the burst windowmaterial at the center measured 101 microns
The above measurements included 15micron total of the material (spray paint)coating both sides of the window
Donald Summers reported a centralthickness of 117 microns (no paint)
Burst test
No fatal events as predicted, a safetyanalysis was submitted before the test forapproval
Pressure was apply to the window/platevolume on increments of 5psig
A leak developed over the elastomer gasketat 133psig
Changed the gasket 2.5mm instead of .8mmdiameter plus vacuum grease
Repeated test, burs pressure at 144psig